Unknown

Dataset Information

Accuracy and clinical relevance of the single-lead Apple Watch electrocardiogram to identify atrial fibrillation.


ABSTRACT:

Background

The Apple Watch (AW) is the first commercially available wearable device with built-in electrocardiogram (ECG) electrodes to perform a single-lead ECG to detect atrial fibrillation (AF).

Methods

Patients with AF who were scheduled for electrical cardioversion (ECV) were included in this study. The AW ECGs were obtained pre-ECV and post-ECV. In case of an unclassified recording, the AW ECG was obtained up to 3 times. The 12-lead ECG was used as the reference standard. Sensitivity, specificity, and kappa coefficient were calculated.

Results

In total, 74 patients were included. Mean age was 67.1 ± 12.3 years and 20.3% were female. In total 65 AF and 64 sinus rhythm measurements were obtained. The first measurement with the AW showed a sensitivity of 93.5% and

SUBMITTER: Pepplinkhuizen S 

PROVIDER: S-EPMC9795256 | biostudies-literature | 2022 Dec

REPOSITORIES: biostudies-literature