Measurement of Physical Activity with Wrist-Worn ActiGraph GT3X+ in Older Women.


ABSTRACT: Higher wear compliance has been seen with wrist placed accelerometers versus hip placed. Performance of wrist placed ActiGraph GT3X+ accelerometer (GT3X+, ActiGraph LLC, Pensacola, FL) in assessing physical activity (PA) remains unclear.

Purpose

This study examined GT3X+'s performance in measuring PA energy expenditure (PAEE) and classifying PA intensity in older women.

Methods

Women [n = 89, age = 65.6 (4.3)] wore GT3X+ and SenseWear Armband Mini (SWAM, BodyMedia Inc. Pittsburgh, PA) for 2 weeks. Concurrently, doubly labeled water (DLW) determined total daily energy expenditure (TDEE). Resting energy expenditure (REE) was determined by Indirect Calorimetry. Data was processed using manufacturer-provided software. Bivariate correlations, Intra Class Correlations, and Bland-Altman plots were performed to evaluate agreement between GT3X+ and criterion measures for sedentary time, light and moderate-to-vigorous PA (determined by SWAM) and PAEE (determined by SWAM and by DLW and REE). Epoch-by-epoch analysis evaluated discrepancy and agreement of PA intensity classification between GT3X+ and SWAM.

Results

For PAEE, GT3X+ showed moderate correlations with criterion measures (r = 0.413, 0.400 with SWAM; r = 0.564, 0.501 with DLW and REE), but Bland-Altman plots showed large variability. When estimating time spent in PA intensity, GT3X+ underestimated sedentary time and overestimated PA intensity compared to SWAM. During epoch-by-epoch analysis, GT3X+ misclassified light intensity PA as moderate-to-vigorous PA 72% of the time. Counts per minute showed strong correlations with criterion measures (r = 0.68, 0.625 for SWAM and DLW and REE respectively).

Conclusion

Current equations and cut points do not provide accurate measures of PA with wrist-worn GT3X+ in older women.

SUBMITTER: Smith MT 

PROVIDER: S-EPMC9797004 | biostudies-literature | 2022

REPOSITORIES: biostudies-literature

altmetric image

Publications

Measurement of Physical Activity with Wrist-Worn ActiGraph GT3X+ in Older Women.

Smith Michal T MT   Kishman Erin E EE   Weaver R Glenn RG   O'Neill Jennifer R JR   Wang Xuewen X  

International journal of exercise science 20221101 7


Higher wear compliance has been seen with wrist placed accelerometers versus hip placed. Performance of wrist placed ActiGraph GT3X+ accelerometer (GT3X+, ActiGraph LLC, Pensacola, FL) in assessing physical activity (PA) remains unclear.<h4>Purpose</h4>This study examined GT3X+'s performance in measuring PA energy expenditure (PAEE) and classifying PA intensity in older women.<h4>Methods</h4>Women [<i>n</i> = 89, age = 65.6 (4.3)] wore GT3X+ and SenseWear Armband Mini (SWAM, BodyMedia Inc. Pitts  ...[more]

Similar Datasets

| S-EPMC7031810 | biostudies-literature
| S-EPMC5561512 | biostudies-literature
| S-EPMC9923693 | biostudies-literature
| S-EPMC6970953 | biostudies-literature
| S-EPMC4993533 | biostudies-literature
| S-EPMC8121263 | biostudies-literature
| S-EPMC8174693 | biostudies-literature
| S-EPMC7234278 | biostudies-literature
| S-EPMC11503582 | biostudies-literature
| S-EPMC7714647 | biostudies-literature