ABSTRACT: Purpose
To develop and perform ex vivo testing for a device designed for semiquantitative determination of intracorneal dissection depth during big bubble (BB) deep anterior lamellar keratoplasty.Methods
A prototype device connected to a syringe and cannula was designed to determine depth of intrastromal placement based on air rebound pressure emitted by a software controlled generator. Ex vivo testing of the device was conducted on human corneas mounted on an artificial anterior chamber in three experiments: (1) cannula purposely introduced at different depths measured with anterior segment optical coherence tomography, (2) cannula introduced as per the BB technique, and (3) simulation of the BB technique guided by the device.Results
A positive pressure differentia