ABSTRACT: Background
In reference dosimetry, radiation quality correction factors are used in order to account for changes in the detector's response among different radiation qualities, improving dosimetric accuracy.Purpose
Reference dosimetry radiation quality corrections factors for the PTW microDiamond were calculated for preclinical X-ray and proton minibeams, and their impact in dosimetric accuracy was evaluated.Methods
A formalism for the calculation of radiation quality correction factors for absolute dosimetry in minibeam fields was developed. Following our formalism, radiation quality correction factors were calculated for the PTW microDiamond detector, using the Monte Carlo method. Models of the detector, and X-ray and proton irradiation platform, were imported int