Unknown

Dataset Information

An in vitro Comparative Assessment of Single-Use Flexible Ureteroscopes Using a Standardized Ureteroscopy Training Model.


ABSTRACT:

Introduction

Perceived benefits like decreased contamination rates and reduced postoperative incidence of complications after urolithiasis surgery have led to increased adoption of single-use flexible ureteroscopes (su-fURS). Using a validated, standardized simulator model with enhanced "fluoroscopic" capabilities, we performed an in vitro comparative assessment of four commercially available models of su-fURS. Both objective and subjective parameters were assessed in this study.

Methods

Two standardized tasks, (1) exploration of the model's kidney collecting system and (2) repositioning of a stone fragment from the upper renal to lower renal pole were assigned to participants, who performed these tasks on all four scopes. Four models of su-fURS (Boston LithoVue, PUSEN PU303

SUBMITTER: So WZ 

PROVIDER: S-EPMC9811416 | biostudies-literature | 2022