Ankle Doppler for Cuffless Ankle Brachial Index Estimation and Peripheral Artery Disease Diagnosis Independent of Diabetes.


ABSTRACT: Ankle brachial pressure index (ABPI) is the first-line test to diagnose peripheral artery disease (PAD). Its adoption in clinical practice is poor and its validity, particularly in diabetes, is limited. We hypothesised that ABPI can be accurately and precisely estimated based on cuffless Doppler waveforms. Retrospective analysis of standard ABPI and handheld Doppler waveform characteristics (n = 200). Prospective analysis of angle-corrected Doppler acceleration index (AccI, n = 148) and standard ABPI with testing of performance to diagnose PAD as assessed with imaging reference standards in consecutive patients. The highest AccI from handheld Doppler at ankle arteries was significantly logarithmically associated with the highest standard ABPI (E[y] = 0.32 ln [1.71 ∗ x + 1], p < 0.001, R2 =