Diagnostic test accuracy of an automated device for the MALDI target preparation for microbial identification.
Ontology highlight
ABSTRACT: The objective of this study was to evaluate the performance of the Copan Colibrí™ against the manual preparation of the MALDI targets. We analyzed 416 (31 different species) non-duplicate strains covering the most important species identified in clinical routine. We also assessed the intra-strain repeatability between the comparable methods. We then analyzed the performance of this new method after implementation in routine on 12,253 aerobic bacterial isolates and yeasts, encompassing a total of 42 different species. Among the 416 strains analyzed, 6.3% (26/416) and 10.8% (45/416) had a score value < 2 when processed by the Colibri™ and manual method, respectively. Only 5.9% (9/152) of the Gram positive rods and cocci had a score values < 2 by the Colibri™ versus 20.4% (31/152) by the manu
SUBMITTER: Cherkaoui A
PROVIDER: S-EPMC9836989 | biostudies-literature | 2023 Feb