Test-retest reliability of the single leg stance on a Lafayette stability platform.
Ontology highlight
ABSTRACT: The validity and reliability of the Lafayette stability platform are well-established for double leg testing. However, no evaluation of single leg (SL) stance on the platform was discovered yet. Therefore, this study aimed to investigate the reliability of conducting the SL stance on the Lafayette platform. Thirty-six healthy and active university students (age 23.2 ± 3.2 years; BMI 21.1 ± 3.1 kg/m2) were tested twice, one week apart (week 1; W1, week 2; W2). They stood on their dominant leg with eyes-open (EO) and eyes-closed (EC) in random order. Three successful trials of 20 seconds each were recorded. The duration during which the platform was maintained within 0° of tilt was referred to as time in balance (TIB). At all-time points, TIB was consistently longer in EO (EOW1: 17.02 ± 1.04