Unknown

Dataset Information

Planar carbon electrodes for real-time quantification of hydrogen sulfide release from cells.


ABSTRACT: A planar electrode system was developed to permit the real-time, selective detection of hydrogen sulfide (H2S) from stimulated cells. Planar carbon electrodes were produced via stencil printing carbon ink through a laser cut vinyl mask. Electrodes were preconditioned using a constant potential amperometry methodology to prevent sensor drift resulting from elemental sulfur adsorption. Modification with a bilaminar coating (electropolymerized ortho-phenylenediamine and a fluorinated xerogel) facilitated high selectivity to H2S. To demonstrate the biological application of this planar sensor system, H2S released from 17β-estradiol-stimulated human umbilical vein endothelial cells (HUVECs) was quantified in situ in real-time. Stimulated HUV

SUBMITTER: Hall JR 

PROVIDER: S-EPMC9850357 | biostudies-literature | 2023 Jan

REPOSITORIES: biostudies-literature