Evaluation of a wrist-worn photoplethysmography monitor for heart rate variability estimation in patients recovering from laparoscopic colon resection.


ABSTRACT: To evaluate the accuracy of heart rate variability (HRV) parameters obtained with a wrist-worn photoplethysmography (PPG) monitor in patients recovering from minimally invasive colon resection to investigate whether PPG has potential in postoperative patient monitoring. 31 patients were monitored for three days or until discharge or reoperation using a wrist-worn PPG monitor (PulseOn, Finland) with a Holter monitor (Faros 360, Bittium Biosignals, Finland) as a reference measurement device. Beat-to-beat intervals (BBI) and HRV information collected by PPG were compared with RR intervals (RRI) and HRV obtained from the ECG reference after removing artefacts and ectopic beats. The beat-to-beat mean error (ME) and mean absolute error (MAE) of good quality heartbeat intervals obtained by wrist