Data Processing of Product Ion Spectra: Quality Improvement by Averaging Multiple Similar Spectra of Small Molecules.


ABSTRACT: In metabolomics studies using high-resolution mass spectrometry (MS), a set of product ion spectra is comprehensively acquired from observed ions using the data-dependent acquisition (DDA) mode of various tandem MS. However, especially for low-intensity signals, it is sometimes difficult to distinguish artifact signals from true fragment ions derived from a precursor ion. Inadequate precision in the measured m/z value is also one of the bottlenecks to narrowing down the candidate compositional formula. In this study, we report that averaging multiple product ion spectra can improve m/z precision as well as the reliability of fragment ions that are observed in such spectra. A graph-based method was applied to cluster a set of similar spectra from multiple DDA data files resulting in creating an averaged product-ion spectrum. The error levels for the m/z values declined following the central limit theorem, which allowed us to reduce the number of candidate compositional formulas. The improved reliability and precision of the averaged spectra will contribute to a more efficient annotation of product ion spectral data.

SUBMITTER: Matsuda F 

PROVIDER: S-EPMC9853114 | biostudies-literature | 2022

REPOSITORIES: biostudies-literature

altmetric image

Publications

Data Processing of Product Ion Spectra: Quality Improvement by Averaging Multiple Similar Spectra of Small Molecules.

Matsuda Fumio F   Komori Shuka S   Yamada Yuki Y   Hara Daiki D   Okahashi Nobuyuki N  

Mass spectrometry (Tokyo, Japan) 20221215 1


In metabolomics studies using high-resolution mass spectrometry (MS), a set of product ion spectra is comprehensively acquired from observed ions using the data-dependent acquisition (DDA) mode of various tandem MS. However, especially for low-intensity signals, it is sometimes difficult to distinguish artifact signals from true fragment ions derived from a precursor ion. Inadequate precision in the measured <i>m</i>/<i>z</i> value is also one of the bottlenecks to narrowing down the candidate c  ...[more]

Similar Datasets

| S-EPMC10713281 | biostudies-literature
| S-EPMC11565486 | biostudies-literature
| S-EPMC3861726 | biostudies-literature
| S-EPMC4662704 | biostudies-literature
| S-EPMC9197348 | biostudies-literature
| S-EPMC3548827 | biostudies-literature
| S-EPMC6657586 | biostudies-literature
| S-EPMC10378377 | biostudies-literature
| S-EPMC6858567 | biostudies-literature
| S-EPMC4570942 | biostudies-literature