Unknown

Dataset Information

Validation of a Temperature-Feedback Controlled Automated Magnetic Hyperthermia Therapy Device.


ABSTRACT: We present in vivo validation of an automated magnetic hyperthermia therapy (MHT) device that uses real-time temperature input measured at the target to control tissue heating. MHT is a thermal therapy that uses heat generated by magnetic materials exposed to an alternating magnetic field. For temperature monitoring, we integrated a commercial fiber optic temperature probe containing four gallium arsenide (GaAs) temperature sensors. The controller device used temperature from the sensors as input to manage power to the magnetic field applicator. We developed a robust, multi-objective, proportional-integral-derivative (PID) algorithm to control the target thermal dose by modulating power delivered to the magnetic field applicator. The magnetic field applicator was a 20 cm diameter Maxwell-t

SUBMITTER: Sharma A 

PROVIDER: S-EPMC9856953 | biostudies-literature | 2023 Jan

REPOSITORIES: biostudies-literature