Unknown

Dataset Information

Analytical performance of 17 commercially available point-of-care tests for CRP to support patient management at lower levels of the health system.


ABSTRACT: Accurate and precise point-of-care (POC) testing for C-reactive protein (CRP) can help support healthcare providers in the clinical management of patients. Here, we compared the analytical performance of 17 commercially available POC CRP tests to enable more decentralized use of the tool. The following CRP tests were evaluated. Eight quantitative tests: QuikRead go (Aidian), INCLIX (Sugentech), Spinit (Biosurfit), LS4000 (Lansionbio), GS 1200 (Gensure Biotech), Standard F200 (SD Biosensor), Epithod 616 (DxGen), IFP-3000 (Xincheng Biological); and nine semi-quantitative tests: Actim CRP (ACTIM), NADAL Dipstick (nal von minden), NADAL cassette (nal von minden), ALLTEST Dipstick (Hangzhou Alltest Biotech), ALLTEST Cassette cut-off 10-40-80 (Hangzhou Alltest Biotech), ALLTEST Cassette cut-off

SUBMITTER: Calarco S 

PROVIDER: S-EPMC9858008 | biostudies-literature | 2023

REPOSITORIES: biostudies-literature