Phantom-based Quality Assurance of a Clinical Dose Accumulation Technique Used in an Online Adaptive Radiation Therapy Platform.


ABSTRACT:

Purpose

This study aimed to develop a routine quality assurance method for a dose accumulation technique provided by a radiation therapy platform for online treatment adaptation.

Methods and materials

Two commonly used phantoms were selected for the dose accumulation QA: Electron density and anthropomorphic pelvis. On a computed tomography (CT) scan of the electron density phantom, 1 target (gross tumor volume [GTV]; insert at 6 o'clock), a subvolume within this target, and 7 organs at risk (OARs; other inserts) were contoured in the treatment planning system (TPS). Two adaptation sessions were performed in which the GTV was recontoured, first at 7 o'clock and then at 5 o'clock. The accumulated dose was exported from the TPS after delivery. Deformable vector fields were also

SUBMITTER: Maraghechi B 

PROVIDER: S-EPMC9860416 | biostudies-literature | 2023 May-Jun