Aerosol Measurement Degradation in Low-Cost Particle Sensors Using Laboratory Calibration and Field Validation.


ABSTRACT: Increasing concern over air pollution has led to the development of low-cost sensors suitable for wide-scale deployment and use by citizen scientists. This project investigated the AirU low-cost particle sensor using two methods: (1) a comparison of pre- and post-deployment calibration equations for 24 devices following use in a field study, and (2) an in-home comparison between 3 AirUs and a reference instrument, the GRIMM 1.109. While differences (and therefore some sensor degradation) were found in the pre- and post-calibration equation comparison, absolute value changes were small and unlikely to affect the quality of results. Comparison tests found that while the AirU did tend to underestimate minimum and overestimate maximum concentrations of particulate matter, ~88% of results fell