Unknown

Dataset Information

Closure device use for common femoral artery antegrade access is higher risk than retrograde access.


ABSTRACT:

Objective

Although the use of closure devices (CD) for femoral artery antegrade access (AA) is not in the instructions for use (IFU) for many devices, AA has been reported to be associated with a lower incidence of access site complications compared to manual compression alone. We hypothesized that CD use for AA would not be associated with a clinically significant increased odds of access site complications compared to CD use for retrograde access (RA).

Methods

This was a retrospective review of the Vascular Quality Initiative from 2010 to 2019 for infrainguinal peripheral vascular interventions with common femoral artery access closed with a CD. Patients who had a cutdown or multiple access sites were excluded. Cases were then stratified into whether access was antegrade o

SUBMITTER: Ramirez JL 

PROVIDER: S-EPMC9869430 | biostudies-literature | 2021 Oct

REPOSITORIES: biostudies-literature