ABSTRACT: Background
This study aimed to investigate how the predicted implant fatigue lifetime is affected by the loss of connector screw preload and the finite element analysis method used to simulate preload.Methods
A dental implant assembly (DI1, Biomet-3i external hex; Zimmer Biomet) was scanned using microcomputed tomography and measured using Mimics software (Materialise) and an optical microscope. Digital replicas were constructed using SolidWorks software (Dassault Systèmes). The material properties were assigned in Abaqus (Dassault Systèmes). An external load was applied at 30° off-axial loading. Eight levels of connector screw preload (range, 0-32 Ncm) were simulated for DI1. This assembly and an additional model (DI2) having a longer and narrower screw were compared regar