Corneal thickness measurements with the RTVue, Casia-2, and Pentacam devices in patients with mild-to-moderate keratoconus: a comparative study.


ABSTRACT:

Background

To compare the characteristics of corneal thickness measurements among the RTVue, Casia-2, and Pentacam in patients with mild-to-moderate keratoconus.

Methods

We recruited 46 eyes of 46 patients diagnosed with mild-to-moderate keratoconus at our hospital between January and March 2022. The central corneal thickness (CCT) and thinnest corneal thickness (TCT) were measured using two optical coherence tomography (OCT) instruments (RTVue and Casia-2) and the more conventional Pentacam. Differences and correlations between the CCTs and TCTs, based on the device and influencing factors, were explored.

Results

The CCTs were highly consistent among the groups (p = 0.434) and correlated with one another (p < 0.001). The TCTs measured by OCTs were thinner than those