In-silico development and assessment of a Kalman filter motor decoder for prosthetic hand control.
Ontology highlight
ABSTRACT: Up to 50% of amputees abandon their prostheses, partly due to rapid degradation of the control systems, which require frequent recalibration. The goal of this study was to develop a Kalman filter-based approach to decoding motoneuron activity to identify movement kinematics and thereby provide stable, long-term, accurate, real-time decoding. The Kalman filter-based decoder was examined via biologically varied datasets generated from a high-fidelity computational model of the spinal motoneuron pool. The estimated movement kinematics controlled a simulated MuJoCo prosthetic hand. This clear-box approach showed successful estimation of hand movements under eight varied physiological conditions with no retraining. The mean correlation coefficient of 0.98 and mean normalized root mean square er
SUBMITTER: Gamal M
PROVIDER: S-EPMC9887730 | biostudies-literature | 2021 May
REPOSITORIES: biostudies-literature