Free-breathing 3D Stack-of-Stars Gradient Echo Sequence in MR-guided Percutaneous Liver Interventions: Evaluation of Workflow and Diagnostic Quality.


ABSTRACT:

Purpose

To evaluate workflow efficiency and diagnostic quality of a free-breathing 3D stack-of-stars gradient echo (Radial GRE) sequence compared to a breath-hold 3D Cartesian gradient echo (Cartesian GRE) sequence for needle position control in MR-guided liver interventions.

Materials and methods

12 MR-guided liver interventions were performed on a 1.5 T Siemens Aera and analyzed retrospectively. 15 series of the Radial GRE sequence were compared to 14 series of the Cartesian GRE sequence regarding the time interval between two consecutive live-scans for needle tracking (Tracking-2-Tracking-Time). The quality of both sequences was compared by the SNR within comparable slices in liver and tumor ROIs. The CNR was calculated by subtraction of the SNR values. Subjective image q