Isolated subcutaneous implantable cardioverter-defibrillator generator displacement causing inappropriate shocks despite preserved lead tip and coil position: a case report.


ABSTRACT:

Background

Subcutaneous implantable cardioverter-defibrillators (S-ICDs) are increasingly used in patients without a pacing indication, and may reduce venous complications, endocarditis and extraction morbidity. Subcutaneous implantable cardioverter-defibrillator displacements may be less obvious than their transvenous counterparts.

Case summary

A 59-year-old man was found to have dilated cardiomyopathy associated with heavy alcohol intake following investigation for a stroke. Despite 2 years of alcohol cessation and optimal drug therapy, his ejection fraction remained severely impaired, and he received an S-ICD using the manufacturer recommended screening and implant method, and by an experienced operator. Ten months later, inappropriate shocks were delivered despite optima