Unknown

Dataset Information

Determination of subpicogram levels of airborne polycyclic aromatic hydrocarbons for personal exposure monitoring assessment.


ABSTRACT: A method based on the use of GC coupled to Q-exactive Orbitrap mass spectrometry (GC-Orbitrap-MS) has been developed for the analysis of polycyclic aromatic hydrocarbons (PAHs) at sub-picogram levels. Outdoor ambient air particulate matter (PM2.5) and standard reference materials (SRM2260a) were analyzed in full scan mode showing low instrumental uncertainties (1-22%) and high linearity over a wide concentration range (0.5 pg and 500 pg/μL). Good reproducibility was obtained compared to the use of conventional single quadrupole GC-MS of PM samples. The quantification limit of the GC-Orbitrap-MS method for full scan analysis of PAHs in outdoor ambient air PM samples was 0.5 pg/μL. This low limit allowed the analysis of PAHs in samples collected with low volumes (< 0.5 m3), such a

SUBMITTER: van Drooge BL 

PROVIDER: S-EPMC9905180 | biostudies-literature | 2023 Feb

REPOSITORIES: biostudies-literature