Unknown

Dataset Information

Suture-based vs. pure plug-based vascular closure devices for VA-ECMO decannulation-A retrospective observational study.


ABSTRACT:

Background

Veno-arterial extracorporeal membrane oxygenation (VA-ECMO) is a valuable treatment option for patients in cardiogenic shock, but complications during decannulation may worsen the overall outcome. Therefore, the aim of this study was to compare the efficacy and safety of suture-based to pure plug-based vascular closure devices for VA-ECMO decannulation.

Methods

In this retrospective study, the procedural outcome of 33 patients with suture-based Perclose ProGlide closure devices was compared to 38 patients with MANTA plug-based closure devices.

Results

Rate of technically correct placement of closure devices was 88% in the suture-based group and 97% in the plug-based group (p = 0.27). There was a significant reduction of severe bleeding events

SUBMITTER: Scherer C 

PROVIDER: S-EPMC9908581 | biostudies-literature | 2023

REPOSITORIES: biostudies-literature