Diagnostic value of real-time four-dimensional transesophageal echocardiography on the implant-related thrombus.


ABSTRACT:

Objectives

This study aims to evaluate the diagnostic value of real-time four-dimensional transesophageal echocardiography (RT4D-TEE) for implant-related thrombus (IRT).

Methods

We collected 1,125 patients with atrial fibrillation from May 2019 to February 2022 in our hospital. All patients accepted transesophageal echocardiography (TEE) examination to exclude any thrombi before the LAAC procedure.

Results

There were 760 patients with LAAC, 66 patients with CIED, and 299 patients without any implantations. A total of 40 patients with an established diagnosis of IRT were further analyzed. The accurate detection rate of IRT by RT4D-TEE was 4.8% (40/826), which was higher than 3.8% (31/826) by 2D-TEE (P = 0.004). No IRT was found on TEE in the rest of the 786 pati