Inter-vendor and inter-observer reliability of diffusion tensor imaging in the musculoskeletal system: a multiscanner MR study.


ABSTRACT:

Background

To evaluate the inter-observer and inter-vendor reliability of diffusion tensor imaging parameters in the musculoskeletal system.

Methods

This prospective study included six healthy volunteers three men (mean age: 42; range: 31-52 years) and three women (mean age: 36; range: 30-44 years). Each subject was scanned using different 3 Tesla magnetic resonance scanners from three different vendors at three different sites bilaterally. First, the intra-class correlation coefficient was used to determine between-observers agreement for overall measurements and clinical sites. Next, between-group comparisons were made through the nonparametric Friedman's test. Finally, the Bland-Altman method was used to determine agreement among the three scanner measurements, comparing them two by two.

Results

A total of 792 measurement were calculated. ICC reported high levels of agreement between the two observers. ICC related to MD, FA, and RD measurements ranged from 0.88 (95% CI 0.85-0.90) to 0.95 (95% CI 0.94-0.96), from 0.85 (95% CI 0.81-0.88) to 0.95 (95% CI 0.93-0.96), and from 0.89 (0.85-0.90) to 0.92 (0.90-0.94). No statistically significant inter-vendor differences were observed. The Bland-Altmann method confirmed a high correlation between parameter values.

Conclusion

An excellent inter-observer and inter-vendor reliability was found in our study.

SUBMITTER: Chianca V 

PROVIDER: S-EPMC9911574 | biostudies-literature | 2023 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Inter-vendor and inter-observer reliability of diffusion tensor imaging in the musculoskeletal system: a multiscanner MR study.

Chianca Vito V   Albano Domenico D   Rizzo Stefania S   Maas Mario M   Sconfienza Luca Maria LM   Del Grande Filippo F  

Insights into imaging 20230209 1


<h4>Background</h4>To evaluate the inter-observer and inter-vendor reliability of diffusion tensor imaging parameters in the musculoskeletal system.<h4>Methods</h4>This prospective study included six healthy volunteers three men (mean age: 42; range: 31-52 years) and three women (mean age: 36; range: 30-44 years). Each subject was scanned using different 3 Tesla magnetic resonance scanners from three different vendors at three different sites bilaterally. First, the intra-class correlation coeff  ...[more]

Similar Datasets

| S-EPMC10543095 | biostudies-literature
| S-EPMC10513221 | biostudies-literature
| S-EPMC5593980 | biostudies-literature
| S-EPMC5953498 | biostudies-literature
| S-EPMC6869410 | biostudies-literature
| S-EPMC9582361 | biostudies-literature
| S-EPMC4990171 | biostudies-literature
| S-EPMC11602640 | biostudies-literature
| S-EPMC4509850 | biostudies-literature
| S-EPMC6593682 | biostudies-literature