Stability of the COVID-19 At-Home Test after Exposure to Extreme Temperatures.


ABSTRACT: To ensure sufficient sensitivity and specificity of lateral flow tests for the detection of SARS-CoV-2 antigen, manufacturers recommend appropriate conditions for storage, including a temperature range. However, there is a high likelihood that kits will be exposed to temperatures outside of this range during transit to some regions. In this prospective study, we evaluated the sensitivity and specificity of the COVID-19 At-Home Test kits (manufactured by SD Biosensor/distributed by Roche) currently being delivered through a US Government program, after exposure to a range of hot and cold temperatures. COVID-19 At-Home Test kits were stored at up to 5 different temperatures: frozen (-4.0°F [-20.0°C]), refrigerated (42.8°F [6.0°C]), room temperature (68.0°F [20.0°C]), warm (98.0°F [36.7°C]),

SUBMITTER: Gick E 

PROVIDER: S-EPMC9927546 | biostudies-literature | 2023 Feb

REPOSITORIES: biostudies-literature