Unknown

Dataset Information

The multi-grip and standard myoelectric hand prosthesis compared: does the multi-grip hand live up to its promise?


ABSTRACT:

Background

Multi-grip myoelectric hand prostheses (MHPs), with five movable and jointed fingers, have been developed to increase functionality. However, literature comparing MHPs with standard myoelectric hand prostheses (SHPs) is limited and inconclusive. To establish whether MHPs increase functionality, we compared MHPs with SHPs on all categories of the International Classification of Functioning, Disability, and Health-model (ICF-model).

Methods

MHP users (N = 14, 64.3% male, mean age = 48.6 years) performed physical measurements (i.e., Refined Clothespin Relocation Test (RCRT), Tray-test, Box and Blocks Test, Southampton Hand Assessment Procedure) with their MHP and an SHP to compare the joint angle coordination and functionality related to the ICF-categories 'Body Func

SUBMITTER: Kerver N 

PROVIDER: S-EPMC9930076 | biostudies-literature | 2023 Feb

REPOSITORIES: biostudies-literature