Diagnostic accuracy of handheld fundus photography: A comparative study of three commercially available cameras.
Ontology highlight
ABSTRACT: The objective of this study was to compare the sensitivity and specificity of handheld fundus cameras in detecting diabetic retinopathy (DR), diabetic macular edema (DME), and macular degeneration. Participants in the study, conducted at Maharaj Nakorn Hospital in Northern Thailand between September 2018 and May 2019, underwent an ophthalmologist examination as well as mydriatic fundus photography with three handheld fundus cameras (iNview, Peek Retina, Pictor Plus). Photographs were graded and adjudicated by masked ophthalmologists. Outcome measures included the sensitivity and specificity of each fundus camera for detecting DR, DME, and macular degeneration, relative to ophthalmologist examination. Fundus photographs of 355 eyes from 185 participants were captured with each of the three
SUBMITTER: Lu L
PROVIDER: S-EPMC9931246 | biostudies-literature | 2022 Nov