The choice of an autocorrelation length in dark-field lung imaging.
Ontology highlight
ABSTRACT: Respiratory diseases are one of the most common causes of death, and their early detection is crucial for prompt treatment. X-ray dark-field radiography (XDFR) is a promising tool to image objects with unresolved micro-structures such as lungs. Using Talbot-Lau XDFR, we imaged inflated porcine lungs together with Polymethylmethacrylat (PMMA) microspheres (in air) of diameter sizes between 20 and 500 [Formula: see text] over an autocorrelation range of 0.8-5.2 [Formula: see text]. The results indicate that the dark-field extinction coefficient of porcine lungs is similar to that of densely-packed PMMA spheres with diameter of [Formula: see text], which is approximately the mean alveolar structure size. We evaluated that, in our case, the autocorrelation length would have to be limited to [F