Long term positional stability of the Argus II retinal prosthesis epiretinal implant.


ABSTRACT:

Background

The Argus II Retinal Prosthesis System (Second Sight Medical Products, Sylmar, California) is an epiretinal prosthesis that serves to provide useful vision to people who are affected by retinal degenerative diseases such as retinitis pigmentosa (RP). The purpose of this study was to analyze postoperative movement of the electrode array.

Methods

Five patients diagnosed with profound retinal dystrophy who have undergone implantation of retinal prosthesis at Stony Brook University Hospital. Fundoscopy was performed at postoperative month 1 (M1), month 3 (M3), month 6 (M6), month 12 (M12), and month 24 (M24) visits. Fundoscopy was extracted and analyzed via NIH ImageJ. Data analysis was completed using IBM SPSS. Various lengths and angles were measured each postoperat