ABSTRACT: Background
Endoscope tips are heated by their electrical and illuminating components. During the procedure, they might get in close or even direct contact with intestinal tissues. Objective
To assess endoscope tip and tissue temperature as well as histopathologic changes of gastrointestinal (GI) tissues when exposed to the heated tip of GI endoscopes. Methods
The endoscope tip temperatures of four GI endoscopes were measured for 30 minutes in a temperature-controlled chamber. The temperature of ex vivo porcine GI tissues was measured for 5-, 15-, and 120-minute exposure to endoscope tips within a climate chamber to control environmental factors (simulation of fever as worst-case). Exposed tissues were histopathologically examined afterward. Control samples included untreated mucosa, tissue samples exposed to endoscope tips for 120 minutes, as well as tissue samples thermally coagulated with a bipolar high-frequency probe. Results
Actual endoscope tip temperatures of 59 to 86°C, dependent on the endoscope type, were measured. After 10 to 15 minutes, the maximum temperatures were reached. Maximum tissue temperatures of 44 to 46°C for 5 and 15 minutes, as well as up to 50°C for 120 minutes, were recorded dependent on tissue and endoscope type. No direct heat-induced histopathologic tissue alterations were observed in the 5- and 15-minute samples. Conclusions
Both clinically relevant and a worst-case control were tested. Even though elevated temperatures were recorded, no heat-related tissue alterations were detected. This overall supports the safety profile of GI endoscopy; however, the study findings are limited by the ex vivo setting (no metabolic tissue alterations accessible, no blood flow) and small sample number.
SUBMITTER: Jager L
PROVIDER: S-EPMC9937900 | biostudies-literature | 2023 Jan
REPOSITORIES: biostudies-literature