ABSTRACT: Introduction
Dual mobility implants have been successful in reducing postoperative hip dislocation but mid-term results of cup migration and polyethylene wear are missing in the literature. Therefore, we measured migration and wear at 5-year follow-up using radiostereometric analysis (RSA). Materials and methods
A cohort of 44 patients (mean age 73, 36 female) with heterogeneous indications for hip arthroplasty but all with a high risk of hip dislocation received total hip replacement (THA) with The Anatomic Dual Mobility X3 monoblock acetabular construct and a highly crosslinked polyethylene liner. RSA images and Oxford Hip Scores were obtained perioperatively and 1, 2, and 5 years postoperatively. Cup migration and polyethylene wear were calculated using RSA. Result