Unknown

Dataset Information

A technical review of bail-out procedures to place Najuta stent-graft into the ascending aorta.


ABSTRACT:

Background

The Najuta stent-graft (Kawasumi Laboratories Inc., Tokyo, Japan) is usually easily advanced to the correct deployment position in the ascending aorta thanks to the pre-curved delivery J-sheath with all fenestrations automatically oriented towards the supra-aortic vessels. Aortic arch anatomy and delivery system stiffness could however represent limitations for proper endograft advancement, especially when the aortic arch bends sharply. The aim of this technical note is to report a series of bail-out procedures that could be useful to overcome the difficulties encountered during the Najuta stent-graft advancement up to the ascending aorta.

Main body

The insertion, positioning and deployment of a Najuta stent-graft requires a through-and-through guidewire technique

SUBMITTER: Berchiolli R 

PROVIDER: S-EPMC9944130 | biostudies-literature | 2023 Feb

REPOSITORIES: biostudies-literature