Wrist-worn device combining PPG and ECG can be reliably used for atrial fibrillation detection in an outpatient setting.


ABSTRACT:

Aims

The aim was to validate the performance of a monitoring system consisting of a wrist-worn device and a data management cloud service intended to be used by medical professionals in detecting atrial fibrillation (AF).

Methods

Thirty adult patients diagnosed with AF alone or AF with concomitant flutter were recruited. Continuous photoplethysmogram (PPG) and intermittent 30 s Lead I electrocardiogram (ECG) recordings were collected over 48 h. The ECG was measured four times a day at prescheduled times, when notified due to irregular rhythm detected by PPG, and when self-initiated based on symptoms. Three-channel Holter ECG was used as the reference.

Results

The subjects recorded a total of 1,415 h of continuous PPG data and 3.8 h of intermittent ECG data over the st