Severe Hypercapnia during Anaesthesia under Mechanical Ventilation in Two Paediatric Patients.


ABSTRACT: A 2-month-old male 1.56 kg Yorkshire terrier (Case No. 1) and a 3-month-old male 2.3 kg Jack Russell Terrier (Case No. 2) were scheduled for ophthalmological surgery under general anaesthesia and neuromuscular blockade. For both patients, volume-controlled ventilation (VCV) was used with set tidal volumes (VT) of 13 mL/kg and 20 mL/kg for cases No. 1 and 2, respectively. The type of ventilator used did not take into account the intrinsic compliance of the breathing system; therefore, a significant part of the delivered VT was wasted in the expansion of the breathing system, and did not reach the patients, causing alveolar hypoventilation. Both cases developed low dynamic compliance (CD), and after a recruitment manoeuvre, EtCO2 of up to 116 mmHg

SUBMITTER: Papastefanou A 

PROVIDER: S-EPMC9951664 | biostudies-literature | 2023 Feb