Thicker polyethylene inserts (≥ 13 mm) increase the risk for early failure after primary cruciate-retaining total knee arthroplasty (TKA): a single-centre study of 7643 TKAs.


ABSTRACT:

Purpose

This study investigates whether thicker (PE) inserts lead to a greater risk for revision after TKA. The differences between the TKA designs of three manufacturers (NexGen, PFC Sigma, Triathlon) are also compared.

Methods

A total of 7643 primary TKA surgeries were included. PE inserts were divided into two groups-"thick PE inserts" with a thickness of 13 mm (mm) or more and "standard PE inserts" with a thickness of less than 13 mm. Three cruciate-retaining (CR) TKA designs (NexGen, PFC Sigma, Triathlon) were included in the study. The differences in failure rates between groups were investigated using Kaplan-Meier survival curves and Cox regression model with hazard ratios (HR). Failure rates were investigated short-term (< 2 years) and long-term (the whole follow-up period). The TKA designs were analysed both together and separately.

Results

During the whole follow-up period, there were 184 (2.4%) aseptic revisions. The thick PE insert group showed an increased risk for revision compared to the standard PE insert group in both short-term (< 2 years; HR 2.0, CI 1.3 to 3.2) and long term (> 2 years; HR 1.6, CI 1.1 to 2.3) follow-up. The highest revision rate was observed in patients who received the Triathlon TKA with a thicker PE insert (HR 2.6, CI 1.2 to 5.7).

Conclusion

The results indicate that thicker PE inserts are associated with increased risk for revision in primary TKA. Further research is required to ascertain whether more conformed PE inserts or constrained knee designs instead of thick CR inserts will ultimately lead to better clinical outcomes.

Level of evidence

III.

SUBMITTER: Rajamaki A 

PROVIDER: S-EPMC9957842 | biostudies-literature | 2023 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Thicker polyethylene inserts (≥ 13 mm) increase the risk for early failure after primary cruciate-retaining total knee arthroplasty (TKA): a single-centre study of 7643 TKAs.

Rajamäki Anni A   Niemeläinen Mika M   Junnila Mika M   Lehtovirta Lari L   Karsikas Mari M   Ponkilainen Ville V   Eskelinen Antti A  

Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA 20221007 3


<h4>Purpose</h4>This study investigates whether thicker (PE) inserts lead to a greater risk for revision after TKA. The differences between the TKA designs of three manufacturers (NexGen, PFC Sigma, Triathlon) are also compared.<h4>Methods</h4>A total of 7643 primary TKA surgeries were included. PE inserts were divided into two groups-"thick PE inserts" with a thickness of 13 mm (mm) or more and "standard PE inserts" with a thickness of less than 13 mm. Three cruciate-retaining (CR) TKA designs  ...[more]

Similar Datasets

| S-EPMC7782324 | biostudies-literature
| S-EPMC2816757 | biostudies-literature
| S-EPMC11867298 | biostudies-literature
| S-EPMC6031546 | biostudies-literature
| S-EPMC11747172 | biostudies-literature
| S-EPMC2565016 | biostudies-other
| S-EPMC2565045 | biostudies-other
| S-EPMC8576761 | biostudies-literature