Thicker polyethylene inserts (≥ 13 mm) increase the risk for early failure after primary cruciate-retaining total knee arthroplasty (TKA): a single-centre study of 7643 TKAs.


ABSTRACT:

Purpose

This study investigates whether thicker (PE) inserts lead to a greater risk for revision after TKA. The differences between the TKA designs of three manufacturers (NexGen, PFC Sigma, Triathlon) are also compared.

Methods

A total of 7643 primary TKA surgeries were included. PE inserts were divided into two groups-"thick PE inserts" with a thickness of 13 mm (mm) or more and "standard PE inserts" with a thickness of less than 13 mm. Three cruciate-retaining (CR) TKA designs (NexGen, PFC Sigma, Triathlon) were included in the study. The differences in failure rates between groups were investigated using Kaplan-Meier survival curves and Cox regression model with hazard ratios (HR). Failure rates were investigated short-term (< 2 years) and long-term (the whole follow-up