Simultaneous assessment of heart and lungs with gated high-pitch ultra-low dose chest CT using artificial intelligence-based calcium scoring.


ABSTRACT:

Purpose

The combined testing for coronary artery and pulmonary diseases is of clinical interest as risk factors are shared. In this study, a novel ECG-gated tin-filtered ultra-low dose chest CT protocol (GCCT) for integrated heart and lung acquisition and the applicability of artificial intelligence (AI)-based coronary artery calcium scoring were assessed.

Methods

In a clinical registry of 10481 patients undergoing heart and lung CT, GCCT was applied in 44 patients on a dual-source CT. Coronary calcium scans (CCS) with 120 kVp, 100 kVp, and tin-filtered 100 kVp (Sn100) of controls, matched with regard to age, sex, and body-mass index, were retrieved from the registry (ntotal=176, 66.5 (59.4-74.0) years, 52 men). Automatic tube current modulation was used in all scans. In 20 patients undergoing GCCT and Sn100 CCS, Agatston scores were measured both semi-automatically by experts and by AI, and classified into six groups (0, <10, <100, <400, <1000, ≥1000).

Results

Effective dose decreased significantly from 120 kVp CCS (0.50 (0.41-0.61) mSv) to 100 kVp CCS (0.34 (0.26-0.37) mSv) to Sn100 CCS (0.14 (0.11-0.17) mSv). GCCT showed higher values (0.28 (0.21-0.32) mSv) than Sn100 CCS but lower than 120 kVp and 100 kVp CCS (all p < 0.05) despite greater scan length. Agatston scores correlated strongly between GCCT and Sn100 CCS in semi-automatic and AI-based measurements (both ρ = 0.98, p < 0.001) resulting in high agreement in Agatston score classification (κ = 0.97, 95% CI 0.92-1.00; κ = 0.89, 95% CI 0.79-0.99). Regarding chest findings, further diagnostic steps were recommended in 28 patients.

Conclusions

GCCT allows for reliable coronary artery disease and lung cancer screening with ultra-low radiation exposure. GCCT-derived Agatston score shows excellent agreement with standard CCS, resulting in equivalent risk stratification.

SUBMITTER: Andre F 

PROVIDER: S-EPMC9958356 | biostudies-literature | 2023

REPOSITORIES: biostudies-literature

altmetric image

Publications

Simultaneous assessment of heart and lungs with gated high-pitch ultra-low dose chest CT using artificial intelligence-based calcium scoring.

Andre Florian F   Seitz Sebastian S   Fortner Philipp P   Allmendinger Thomas T   Sommer André A   Brado Matthias M   Sokiranski Roman R   Fink Joana J   Kauczor Hans-Ulrich HU   Heussel Claus P CP   Herth Felix F   Frey Norbert N   Görich Johannes J   Buss Sebastian J SJ  

European journal of radiology open 20230215


<h4>Purpose</h4>The combined testing for coronary artery and pulmonary diseases is of clinical interest as risk factors are shared. In this study, a novel ECG-gated tin-filtered ultra-low dose chest CT protocol (GCCT) for integrated heart and lung acquisition and the applicability of artificial intelligence (AI)-based coronary artery calcium scoring were assessed.<h4>Methods</h4>In a clinical registry of 10481 patients undergoing heart and lung CT, GCCT was applied in 44 patients on a dual-sourc  ...[more]

Similar Datasets

| S-EPMC11320525 | biostudies-literature
| S-EPMC7656362 | biostudies-literature
| S-EPMC11356370 | biostudies-literature
| S-EPMC7429815 | biostudies-literature
| S-EPMC9406755 | biostudies-literature
| S-EPMC10607407 | biostudies-literature
| S-EPMC9530804 | biostudies-literature
| S-EPMC8760133 | biostudies-literature
| S-EPMC7083991 | biostudies-literature
| S-EPMC6861262 | biostudies-literature