A Hybrid Stainless-Steel SPME Microneedle Electrode Sensor for Dual Electrochemical and GC-MS Analysis
Ontology highlight
ABSTRACT: A mechanically robust in-tube stainless steel microneedle solid phase microextraction (SPME) platform for dual electrochemical and chromatographic detection has been demonstrated. The SPME microneedle was fabricated by layer-by-layer (LbL) in-tube coating, consisting of carbon nanotube (CNT)/cellulose nanocrystal (CNC) film layered with an electrically conductive polyaniline (PANI) hydrogel layer (PANI@CNT/CNC SPME microneedle (MN)). The PANI@CNT/CNC SPME MN showed effective analysis of caffeine by GC-MS with an LOD of 26 mg/L and excellent precision across the dynamic range. Additionally, the PANI@CNT/CNC SPME MN demonstrated a 67% increase in sensitivity compared to a commercial SPME fiber, while being highly robust for repeated use without loss in performance. For electrochemical detect
SUBMITTER:
PROVIDER: S-EPMC9963686 | biostudies-literature | 2023 Feb
REPOSITORIES: biostudies-literature
ACCESS DATA