Evaluation of the Long-Term Clinical Results of 3 Patients Implanted with the Argus II Retinal Prosthesis.


ABSTRACT: This study presents the long-term clinical results of Argus II retinal prosthesis implantation in eyes with light perception and projection in 3 patients with end-stage retinitis pigmentosa. No conjunctival erosion, hypotony, or implant displacement was observed during postoperative follow-up. The electrical threshold values were lower in the macular region and higher close to the tack fixation region and peripherally. Optical coherence tomography scans showed fibrosis and retinoschisis formation at the retina-implant interface in two cases. This was attributed to mechanical and electrical effects on the tissue due to the active daily use of the system and the electrodes' proximity to the retina. The patients were able to integrate the system into their daily lives and perform activities t