Effects of spatial and temporal resolution on cardiovascular magnetic resonance feature tracking measurements using a simple realistic numerical phantom.


ABSTRACT:

Objectives

To develop a single-slice numerical phantom with known myocardial motion, at several temporal and in-plane spatial resolutions, for testing and comparison of Cardiovascular Magnetic Resonance (CMR) feature tracking (FT) software.

Methods

The phantom was developed based on CMR acquisitions of one volunteer (acquired cine, tagging cine, T1 map, T2 map, proton density weighted image). The numerical MRI simulator JEMRIS was used, and the phantom was generated at several in-plane spatial resolutions (1.4 × 1.4 mm2 to 3.0 × 3.0 mm2) and temporal resolutions (20 to 40 cardiac phases). Two feature tracking software packages were tested: Medical Image Tracking Toolbox (MITT) and two versions of cvi42 (v5.3.8 and v5.13.7). The effect of resolution on s