Comparative dosimetric analysis of volumetric modulated arc therapy based craniospinal irradiation plans between Halcyon ring gantry and TrueBeam C-arm linear accelerator.
Ontology highlight
ABSTRACT: This study evaluates the volumetric modulated arc therapy (VMAT) dosimetric comparison between Halcyon ring gantry and TrueBeam c-arm linear accelerators for craniospinal irradiation (CSI) of the neuro-axis. 25 patients, who received treatment for medulloblastoma and primitive neuro-ectodermal tumors between 2018 and 2021, were planned for VMAT in True Beam (TB), and Halcyon (HAL) linear accelerators using 6 MV unflattened (FFF) photon beams (HALFFF and TBFFF). Dose-volume statistics for the target and organs at risk (OARs) and the total number of monitoring units (MUs) in the treatment plans were compared which included dose received by 95% PTV volume (V95%), volume receiving ≥ 107% dose, homogeneity index (HI), conformity index (PI), MU and dose spillage (D10%, D30%, D50%, D70%, D90%). I