ABSTRACT: Purpose
To describe two new stereoacuity tests: the eRDS v6 stereotest, a global dynamic random dot stereogram (dRDS) test, and the Vivid Vision Stereo Test version 2 (VV), a local or "contour" stereotest for virtual reality (VR) headsets; and to evaluate the tests' reliability, validity compared to a dRDS standard, and learning effects.Methods
Sixty-four subjects passed a battery of stereotests, including perceiving depth from RDS. Validity was evaluated relative to a tablet-based dRDS reference test, ASTEROID. Reliability and learning effects were assessed over six sessions.Results
eRDS v6 was effective at measuring small thresholds (<10 arcsec) and had a moderate correlation (0.48) with ASTEROID. Across the six sessions, test-retest reliability was good, varying