The Accuracy of Wrist-Worn Photoplethysmogram-Measured Heart and Respiratory Rates in Abdominal Surgery Patients: Observational Prospective Clinical Validation Study.


ABSTRACT:

Background

Postoperative deterioration is often preceded by abnormal vital parameters. Therefore, vital parameters of postoperative patients are routinely measured by nursing staff. Wrist-worn sensors could potentially provide an alternative tool for the measurement of vital parameters in low-acuity settings. These devices would allow more frequent or even continuous measurements of vital parameters without relying on time-consuming manual measurements, provided their accuracy in this clinical population is established.

Objective

This study aimed to assess the accuracy of heart rate (HR) and respiratory rate (RR) measures obtained via a wearable photoplethysmography (PPG) wristband in a cohort of postoperative patients.

Methods

The accuracy of the wrist-worn PPG sensor was assessed in 62 post-abdominal surgery patients (mean age 55, SD 15 years; median BMI 34, IQR 25-40 kg/m2). The wearable obtained HR and RR measurements were compared to those of the reference monitor in the postanesthesia or intensive care unit. Bland-Altman and Clarke error grid analyses were performed to determine agreement and clinical accuracy.

Results

Data were collected for a median of 1.2 hours per patient. With a coverage of 94% for HR and 34% for RR, the device was able to provide accurate measurements for the large majority of the measurements as 98% and 93% of the measurements were within 5 bpm or 3 rpm of the reference signal. Additionally, 100% of the HR and 98% of the RR measurements were clinically acceptable on Clarke error grid analysis.

Conclusions

The wrist-worn PPG device is able to provide measurements of HR and RR that can be seen as sufficiently accurate for clinical applications. Considering the coverage, the device was able to continuously monitor HR and report RR when measurements of sufficient quality were obtained.

Trial registration

ClinicalTrials.gov NCT03923127; https://www.clinicaltrials.gov/ct2/show/NCT03923127.

SUBMITTER: van der Stam JA 

PROVIDER: S-EPMC9989911 | biostudies-literature | 2023 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

The Accuracy of Wrist-Worn Photoplethysmogram-Measured Heart and Respiratory Rates in Abdominal Surgery Patients: Observational Prospective Clinical Validation Study.

van der Stam Jonna A JA   Mestrom Eveline H J EHJ   Scheerhoorn Jai J   Jacobs Fleur E N B FENB   Nienhuijs Simon S   Boer Arjen-Kars AK   van Riel Natal A W NAW   de Morree Helma M HM   Bonomi Alberto G AG   Scharnhorst Volkher V   Bouwman R Arthur RA  

JMIR perioperative medicine 20230220


<h4>Background</h4>Postoperative deterioration is often preceded by abnormal vital parameters. Therefore, vital parameters of postoperative patients are routinely measured by nursing staff. Wrist-worn sensors could potentially provide an alternative tool for the measurement of vital parameters in low-acuity settings. These devices would allow more frequent or even continuous measurements of vital parameters without relying on time-consuming manual measurements, provided their accuracy in this cl  ...[more]

Similar Datasets

| S-EPMC5050383 | biostudies-literature
| S-EPMC6913722 | biostudies-literature
| S-EPMC5564399 | biostudies-literature
| S-EPMC5984239 | biostudies-literature
| S-EPMC6550265 | biostudies-literature
| S-EPMC5491979 | biostudies-literature
| S-EPMC10747116 | biostudies-literature
| S-EPMC9923693 | biostudies-literature
| S-EPMC9624389 | biostudies-literature
| S-EPMC7794504 | biostudies-literature