ABSTRACT: Aims
Glucose and insulin metabolism are altered in hemodialysis patients, and diabetes management is difficult in these patients. We aimed to validate flash glucose monitoring (FGM) in hemodialysis patients with and without diabetes mellitus as an attractive option for glucose monitoring not requiring regular self-punctures.Methods
We measured interstitial glucose using a FreeStyle Libre device in eight hemodialysis patients with and seven without diabetes mellitus over 14 days and compared the results to simultaneously performed self-monitoring of capillary blood glucose (SMBG).Results
In 720 paired measurements, mean flash glucose values were significantly lower than self-measured capillary values (6.17±2.52 vs. 7.15±2.41 mmol/L, p=1.3 E-86). Overall, the mean abs