Study of the effect of distance and misalignment between magnetically coupled coils for wireless power transfer in intraocular pressure measurement.


ABSTRACT: An analysis of the effect of distance and alignment between two magnetically coupled coils for wireless power transfer in intraocular pressure measurement is presented. For measurement purposes, a system was fabricated consisting of an external device, which is a Maxwell-Wien bridge circuit variation, in charge of transferring energy to a biomedical implant and reading data from it. The biomedical implant is an RLC tank circuit, encapsulated by a polyimide coating. Power transfer was done by magnetic induction coupling method, by placing one of the inductors of the Maxwell-Wien bridge circuit and the inductor of the implant in close proximity. The Maxwell-Wien bridge circuit was biased with a 10 MHz sinusoidal signal. The analysis presented in this paper proves that wireless transmission of power for intraocular pressure measurement is feasible with the measurement system proposed. In order to have a proper inductive coupling link, special care must be taken when placing the two coils in proximity to avoid misalignment between them.

SUBMITTER: Rendon-Nava AE 

PROVIDER: S-EPMC4109074 | biostudies-other | 2014

REPOSITORIES: biostudies-other

Similar Datasets

| S-EPMC6940365 | biostudies-literature
| S-EPMC10356933 | biostudies-literature
| S-EPMC7034622 | biostudies-literature
| S-EPMC6594360 | biostudies-literature
| S-EPMC5818941 | biostudies-literature
| S-EPMC5710491 | biostudies-literature
| S-EPMC8878591 | biostudies-literature
| S-EPMC9114010 | biostudies-literature
| S-EPMC7875381 | biostudies-literature
| S-EPMC9649789 | biostudies-literature