The British journal of radiology 20150326 1050
<h4>Objective</h4>A dedicated extremity cone beam CT (CBCT) was introduced recently, and is rapidly becoming an attractive modality for extremity imaging. This study aimed to evaluate the effectiveness of a curtain-shaped lead shielding in reducing the exposure of patients to scattered radiation in dedicated extremity CBCT.<h4>Methods</h4>A dedicated extremity CBCT scanner was used. The lead shielding curtain was 42 × 60 cm with 0.5-mm lead equivalent. Scattered radiation dose from CBCT was meas ...[more]