20230301_LCMSMS_pairedwithLCICPMS_scopulariopsis


ABSTRACT: High resolution LC-MS/MS data for pooled biological replicates (N=3) was collected on a Bruker compact qTOF in positive mode with the detection window set from 100 - 2000 Da on an ACE excel 1.7 super C18 2.1 x 100 mm UPLC column with a flow rate of 0.35 mL/min. Agilent ESI TOF mix calibrant was injected via divert valve from minute 0.04 to minute 0.30 during the LC run. Solvent A was MilliQ water with 5 mM ammonium formate adjusted to pH 6.5 using ammonium hydroxide. Solvent B was 95.5 ACN:MilliQ water with 5 mM ammonium formate adjusted to pH 6.5 using ammonium hydroxide. Solvent was held at 0% B for 2 minutes and then a gradient of 0 - 100% B over 15 minutes was run and brought back down to 0 % at 15.2 minutes. For each sample, 20 uL of a 1 mg/mL solution was injected. The ESI conditions were set with a capillary voltage at 4.5 kV. For MS/MS, dynamic exclusion was used, and the top nine precursor ions from each MS1 scan were subjected to collision energies scaled according to mass and charge state for a total of nine data dependent MS/MS events per MS1.

INSTRUMENT(S): compact

ORGANISM(S): Scopulariopsis Sp. 165-5 Scopulariopsis Sp. Jb370