Magnetic Resonance Safety Evaluation of a Novel Alumina Matrix Composite Ceramic Knee and Image Artifact Comparison to a Metal Knee Implant of Analogous Design.


ABSTRACT:

Background

Image artifacts caused by metal knee implants in 1.5T and 3T magnetic resonance imaging (MRI) systems complicate imaging-based diagnosis of the peri-implant region after total knee arthroplasty. Alternatively, metal-free knee prostheses could effectively minimize MRI safety hazards and offer the potential for higher quality diagnostic images.

Methods

A novel knee arthroplasty device composed of BIOLOX delta, an alumina matrix composite (AMC) ceramic, was tested in a magnetic resonance (MR) environment. American Society for Testing and Materials test methods were used for evaluating magnetically induced displacement force, magnetically induced torque, radiofrequency-induced heating, and MR image artifact.

Results

Magnetically induced displacement force and magnetically induced torque results of the AMC ceramic knee indicated that these effects do not pose a known risk in a clinical MR environment, as assessed in a 3T magnetic field. Moreover, minimal radiofrequency-induced heating of the device was observed. In addition, the AMC ceramic knee demonstrated minimal MR image artifacts (7 mm) in comparison to a cobalt-chromium knee (88 mm). The extremely low magnetic susceptibility of AMC (2 ppm) underlines that it is a nonmetallic and nonmagnetic material well suited for the manufacturing of MR Safe orthopaedic implants.

Conclusions

The AMC ceramic knee is a novel metal-free total knee arthroplasty device that can be regarded as MR Safe, as suggested by the absence of hazards from the exposure of this implant to a MR environment. The AMC ceramic knee presents the advantage of being scanned with superior imaging results in 3T MRI systems compared to alternative metal implants on the market.

SUBMITTER: Modinger Y 

PROVIDER: S-EPMC10374871 | biostudies-literature | 2023 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Magnetic Resonance Safety Evaluation of a Novel Alumina Matrix Composite Ceramic Knee and Image Artifact Comparison to a Metal Knee Implant of Analogous Design.

Mödinger Yvonne Y   Anttila Eric D ED   Baker Grant M GM   Gross David C DC   Porporati Alessandro A AA  

Arthroplasty today 20230723


<h4>Background</h4>Image artifacts caused by metal knee implants in 1.5T and 3T magnetic resonance imaging (MRI) systems complicate imaging-based diagnosis of the peri-implant region after total knee arthroplasty. Alternatively, metal-free knee prostheses could effectively minimize MRI safety hazards and offer the potential for higher quality diagnostic images.<h4>Methods</h4>A novel knee arthroplasty device composed of BIOLOX <i>delta</i>, an alumina matrix composite (AMC) ceramic, was tested i  ...[more]

Similar Datasets

| S-EPMC7878514 | biostudies-literature
| S-EPMC6196667 | biostudies-literature
| S-EPMC6965657 | biostudies-literature
| S-EPMC7263301 | biostudies-literature
| S-EPMC5045460 | biostudies-literature
| S-EPMC6366181 | biostudies-literature
| S-EPMC8154748 | biostudies-literature
| S-EPMC7583733 | biostudies-literature
| S-EPMC9445226 | biostudies-literature
| S-EPMC4046129 | biostudies-literature