Unknown

Dataset Information

Bailout clipping of a leaflet perforation during mitral transcatheter edge-to-edge repair using a larger clip size: a case report.


ABSTRACT:

Background

Leaflet tear and perforation are serious complications of transcatheter edge-to-edge repair (TEER) using the MitraClip system for severe mitral regurgitation (MR). However, no optimal bailout strategy has been established.

Case summary

An 80-year-old woman developed heart failure due to acute ischaemic severe MR after primary percutaneous coronary intervention. Given the requirement for inotropic drugs and an intra-aortic balloon pump to stabilize her heart failure, we decided to perform TEER using the MitraClip G4 system. The NTW clip was selected considering the limited posterior leaflet, wide central MR, and delivery to the central A2/P2. The leaflet was successfully grasped, and the MR jet dramatically decreased while the clip was closed. However, the MR jet s

SUBMITTER: Nishiura N 

PROVIDER: S-EPMC10500417 | biostudies-literature | 2023 Sep

REPOSITORIES: biostudies-literature