Visual Field Evaluation Using Zippy Adaptive Threshold Algorithm (ZATA) Standard and ZATA Fast in Patients With Glaucoma and Healthy Individuals.


ABSTRACT:

Purpose

To evaluate visual fields obtained with Zippy Adaptive Threshold Algorithm (ZATA) Standard and ZATA Fast from patients with glaucoma and healthy individuals.

Methods

Fifty-five patients with glaucoma (median mean deviation [MD], -7.6 dB; interquartile range [IQR], -15.3 to -2.6 dB) and 22 healthy participants (median MD, -0.6 dB; IQR, -1.7 to 0.2 dB) performed ZATA Standard and ZATA Fast tests on a Henson 9000 perimeter and Swedish Interactive Thresholding Algorithm (SITA) Standard and SITA Fast tests on a Humphrey Field Analyzer. Tests were repeated within 90 days (median, 14 days; range, 7-26 days) to evaluate the test-retest variability.

Results

The mean difference between the MD of the ZATA Standard and SITA Standard tests was 1.7 dB (95% confidence interv